If the DAC UNIVERSAL shows an error in the display, it is important to find out exactly, what this error means and what

action to take. Below you will find a list of the error codes.

If the DAC UNIVERSAL displays an error, you cannot be sure that a complete cycle has been carried out and therefore the

instruments are not sterile – so you must start a complete new cycle.

If an error occurs during sterilization, the display will, when the lid opens, show that an error has occurred during

sterilization. In order to clear this error and start a new complete cycle — you need to press the «M»-button together with the

«C»-button (This will also be indicated in the display)

If the error light is flashing without giving an alarm signal, it is because the machine has reached 3000 cycles or it has

been 24 months since the last service. Contact your supplier in order to do a biennial service on the DAC UNIVERSAL

autoclave. Meanwhile the machine can still be used if „only‟ the error light is flashing and no other error codes are visual in

the display.

In case of an error, an alarm will sound and the error will be displayed visually.

Please note the error code before resetting the alarm. Pressing C resets the alarm.

When pressing C the lid will for approx. 2 seconds raise and then stop, until the temperature is below 104C/219.2F,

following which it will open completely.

It is, however, possible to force the lid to open completely by pressing C repeatedly. For every press on C, the lid will

continue opening for 2 seconds. We do not recommend this feature to be used.

NB: Do not open if the temperature is more than 104C/219.2F (C indicated in the display), hot steam might come out of

the chamber.

Error

64

65

66

67

68

70

71

72

73

74

75

76

77

78

79

81

82

83

84

34

DAC UNIVERSAL – USER MANUAL

729050 — Version 16 S

Description

The safety system has been activated  Press C  Restart the autoclave.

The safety system is defective  Please contact your supplier.

The temperature during sterilization is too high  Please contact your supplier.

The time of sterilization has been less than 3 minutes The instruments should

be sterilised again  Please contact your supplier if this error appears more than

once.

The pressure is too high during pressure test  Please contact your supplier

The pressure level does not decrease fast enough after pressure test  Empty

the chamber for water  Clean or change the filter  Check that the drain tube is

not blocked or broken (note that the drain tube must be of heat resistant material,

e.g. PTFE)  Empty the drain tank if full  Please contact your supplier.

The autoclave was blocked due to use of bad water  Control the level of micro-

siemens in the water  Empty the watertank  Fill in fresh water (below 3S)

The closing system was blocked on the way up  Remove possible blocking for

the lift  Please contact your supplier

The sterilization temperature has not been reached fast enough before the first

back-flush  Pour in water in case of low level  Please contact your supplier

The right temperature for building up pressure for sterilization has not been

reached fast enough  Pour in water if low level  If the autoclave is placed in a

cold spot remove it to surroundings at room temperature  Press C to reset and

then restart  Please contact your supplier

The right sterilization temperature has not been reached fast enough  Pour in

water if low level  If the autoclave is placed in a cold spot move it to a

surrounding at room temperature  Press C to reset and then restart  Please

contact your supplier

Valve error  Please contact your supplier or try with only 5 instruments on the lid.

The autoclave does not lubricate  Please contact your supplier

Valve error  Please contact your supplier or try with only 5 instruments on the lid.

Valve error  Please contact your supplier or try with only 5 instruments on the lid.

Low level in either the water tank or the oil bottle  Pour water into the water tank

or change the oil bottle  Please contact your supplier

The temperature in the steam generator is too high  Please contact your

supplier

Temperature in chamber is too high  Please contact your supplier

The sterilization temperature has not been maintained during sterilization  Pour

in water, if needed  Please contact your supplier

60 53 362 D 3472

D3472.201.01.09.02

Sirona DAC Universal Operating Instructions Manual

  • Contents

  • Table of Contents

  • Bookmarks

Quick Links

MRKOMNT

kÉï=~ë=çÑW=

a^`=råáîÉêë~ä

qÉñí=aáëéä~ó=J=sÉêëáçå=OR

léÉê~íáåÖ=fåëíêìÅíáçåë

båÖäáëÜ

Cover page

=

loading

Related Manuals for Sirona DAC Universal

Summary of Contents for Sirona DAC Universal

  • Page 1
    MRKOMNT kÉï=~ë=çÑW= a^`=råáîÉêë~ä qÉñí=aáëéä~ó=J=sÉêëáçå=OR léÉê~íáåÖ=fåëíêìÅíáçåë båÖäáëÜ Cover page…
  • Page 2
    Operating Instructions DAC Universal Dear Customer, Thank you for purchasing a DAC Universal. The DAC Universal cleans, lubricates and disinfects / sterilizes 6 instruments automatically. Solid instruments can also be cleaned and sterilized. DAC Universal saves you time-consuming manual reprocessing and reliably protects you, your team and your patients from cross-contamination.
  • Page 3
    Sirona Dental A/S Table of Contents Operating Instructions DAC Universal Table of Contents General Information………………..Contact information ………………General notes………………..Electronic «User E-Guide» …………….Other valid documents …………….. Structure of the document …………….1.5.1 Identification of the danger levels………… 1.5.2 Formats and symbols used ………….
  • Page 4
    Table of Contents Sirona Dental A/S Operating Instructions DAC Universal Presentation of the sterilization process …………Electromagnetic compatibility……………. 3.6.1 Electromagnetic emission …………..3.6.2 Interference immunity…………… 3.6.3 Working clearances…………….Safety devices ………………..Transport ………………….. Installation …………………… Requirements of the location / Ambient conditions ……..
  • Page 5
    Sirona Dental A/S Table of Contents Operating Instructions DAC Universal 5.2.5 Drain the chamber…………….Reprocessing with the DAC Universal …………5.3.1 Switching ON/OFF/Standby mode ……….5.3.2 Reprocessing with the Standard lid……….5.3.2.1 Start cycle on the Standard lid………. 5.3.3 Reprocessing with the Basket lid …………
  • Page 6: General Information

    If you reach an impasse despite having thoroughly studied the operating instructions, please contact your dental depot. Electronic «User E-Guide» An electronic «User E-guide» is included with the DAC Universal containing all operating instructions and important technical information. Other valid documents DAC Universal includes other separate documents.

  • Page 7: Structure Of The Document

    Sirona Dental A/S 1 General Information Operating Instructions DAC Universal 1.5 Structure of the document Structure of the document 1.5.1 Identification of the danger levels To prevent personal injury and material damage, please observe the warning and safety information provided in the present operating instructions.

  • Page 8: Warranty And Liability

    Universal) was subjected to a thorough quality test before it left the factory. A warranty period of 12 months is given for DAC Universal starting on the day it is delivered. The warranty covers production and material defects and includes both replacement parts and work.

  • Page 9: Contraindications

    Dental handpieces are defined as rotating instruments split up into the categories straight handpieces, contra angles and turbines. The DAC Universal is intended to clean and steam sterilize solid dental instruments. The DAC Universal is intended to clean and thermally disinfect ultrasonic handpieces.

  • Page 10: Scope Of Delivery

    1 General Information Sirona Dental A/S 1.9 Scope of delivery Operating Instructions DAC Universal Scope of delivery The following components are included in the scope of delivery of the DAC Universal: ● Standard lid without adapters ● Indicator holder, Standard lid ● 1 bottle of NitramOil #2 lubrication concentrate ●…

  • Page 11: Safety Instructions

    2.1 Electrostatic discharge Safety instructions IMPORTANT This product is intended for use by qualified healthcare professionals only. Before operating the DAC Universal, carefully read and follow these instructions and save them for future reference. Observe all cautions. Electrostatic discharge Protective measures Electrostatic discharge (abbreviated: ESD –…

  • Page 12
    2 Safety instructions Sirona Dental A/S 2.1 Electrostatic discharge Operating Instructions DAC Universal An electrostatic discharge is: ● perceptible at 3,000 V or higher ● audible at 5,000 V or higher (cracking, crackling) ● visible at 10,000 V or higher (arc-over) The transient currents resulting from these discharges have a magnitude of over 10 amps.
  • Page 13: Caution Symbols On The Device

    Caution symbols on the device Meaning: Warning of danger This symbol is found on the back of the DAC Universal next to: ● The COM PORT: External computer hardware that is connected to this serial interface must be labeled «The connected device must comply with the UL 1950 and IEC/EN 60950 standards and must be connected only to safety extra low voltage (SELV) circuits.»…

  • Page 14: Flammable Materials

    Universal when it opens, as steam will emerge from the device. CAUTION Risk of burns! Even when the lid of the DAC Universal is partly or completely open, the instruments are still hot. Allow the instruments that become very hot during the process to cool down.

  • Page 15: Instruction Of Practice Personnel

    DAC Universal. Installation site CAUTION Risk of crushing! The total height of the open DAC Universal is 53 cm. To prevent injuries, a minimum distance of 70 cm must be maintained between the counter and overhead cabinet. Maintenance…

  • Page 16: Product Description

    These instruments are referred to as Flex lid instruments. The number and placement of the different adapters is optional. Exception: The Flex lid can contain a maximum of 3 scaler tips of Sirona and / or Satelec simultaneously, which should not be placed next to one another.

  • Page 17
    20 to 30 minutes. To ensure trouble-free operation, accumulated operating time of rotating handpieces treated in DAC Universal shall not exceed 10 minutes. External cleaning The instruments are first cleaned at least 3 times with unheated water in a pulse-wash procedure (multicyclic cleaning method).
  • Page 18
    3.1.2 Functions with the Basket lid 5 solid instruments, instruments without cavities, can be reprocessed in the DAC Universal with the optional Basket lid. WARNING According to the European EN 13060 standard, non-solid instruments or instruments with cavities may not be sterile after a cycle in the Basket lid.
  • Page 19
    Sirona Dental A/S 3 Product description Operating Instructions DAC Universal 3.1 Functions See «NitraClean cleaning tablet [ → 104]» for more information on NitraClean cleaning tablets. Sterilization After external cleaning, the chamber is heated to 134 °C (273 °F) and the instruments are sterilized. Air is extracted from the chamber and the chamber is filled with pressurized, saturated steam.
  • Page 20
    3 Product description Sirona Dental A/S 3.1 Functions Operating Instructions DAC Universal Internal cleaning Internal cleaning ensures that the channels of the instruments are rinsed thoroughly. For the internal cleaning of the instruments, water is forced through the channels using compressed air. The channels are rinsed with water 12 times during a cleaning cycle.
  • Page 21
    Sirona Dental A/S 3 Product description Operating Instructions DAC Universal 3.2 Controls and functional elements Find enclosed the characteristic values of the Flex lid. Flex lid Cleaning Disinfection max. water consumption 470 ml 260 ml max. time 645 s 190 s max. temperature 75 °C (167 °F)
  • Page 22
    3.2 Controls and functional elements Operating Instructions DAC Universal 3.2.2 The following lids can be used for reprocessing in the DAC Universal: ● Standard lid for reprocessing unwrapped rotating handpieces. See «Functions with the Standard lid [ → 16]» for functions and «Reprocessing with the Standard lid [ →…
  • Page 23
    This key can be used for Check & Clean, see «Cleaning and maintenance [ → 77]». G: «Start»-Key This key switches the DAC Universal on and starts it. H: «Lubricate»-Key When lubrication is activated, a blue LED control lamp lights up. Pressing the «Lubricate»- Key when it is switched on activates or deactivates lubrication.
  • Page 24
    See also «Installation of connections [ → 39]». 3.2.5 Safety ring The DAC Universal is equipped with a safety ring. This serves to trigger the process and prevents the user from pinching the fingers when closing the lid. If there is an object between the chamber and the lid when the lid closes, the cycle is stopped immediately.
  • Page 25
    NOTICE The water should be poured carefully through the sieve into the tank to avoid spilling water into the DAC Universal, which may result in water on the table or on the electrical parts of the DAC Universal, which can cause a short circuit.
  • Page 26
    3.3 Programs and program selection Operating Instructions DAC Universal Programs and program selection The DAC Universal has 2 different programs that can be selected as needed. They are set using the menu system, see «Menu system. [ → 59]». Factory default is the standard program (134 °C / 273 °F, sterilization time 3 minutes).
  • Page 27: Technical Data

    Type: Not applicable Type BF applied part: Not applicable Safety class: Never use the DAC Universal in the presence of flammable anesthetics or gases. Cycle length: 18 – 33 minutes Operating conditions: Temperature: 16 ℃ – 26 ℃ (61 ℉ –79 ℉ ) The DAC Universal must not be stored at temperatures below 0 °C (32 °F), while in operating modus.

  • Page 28
    3 Product description Sirona Dental A/S 3.4 Technical data Operating Instructions DAC Universal The DAC Universal complies with the requirements of the RoHS Directive 2011/65/EU. Symbols on the DAC Universal packaging/external carton This way up Keep away from rain Handle with care Use no hooks Temperature limit: -40 °C to +65 °C…
  • Page 29: Presentation Of The Sterilization Process

    Sirona Dental A/S 3 Product description Operating Instructions DAC Universal 3.5 Presentation of the sterilization process Presentation of the sterilization process The following graphs show an overview of the process. This process overview shows the different temperature and pressure levels over time.

  • Page 30
    3 Product description Sirona Dental A/S 3.5 Presentation of the sterilization process Operating Instructions DAC Universal Prion program The following graph shows temperature (°C) and pressure (bar) during the prion program, with standard settings, holding time 18 min and a total cycle time of 33 min:…
  • Page 31
    Sirona Dental A/S 3 Product description Operating Instructions DAC Universal 3.5 Presentation of the sterilization process Basket lid The following graph shows temperature (°C) and pressure (bar) during the standard program with the Basket lid, with standard settings, holding time 3 min and a total cycle time of 18 min:…
  • Page 32
    3 Product description Sirona Dental A/S 3.5 Presentation of the sterilization process Operating Instructions DAC Universal Flex lid The following graph shows temperature (°C) and pressure (bar) during the standard program with the Flex lid, with standard settings, holding time 3 min and a total cycle time of 21 min:…
  • Page 33
    The table below gives an overview of what is performed during which programs mentioned in «Functions [ → 16]“. Type test according to applicable standards Solid instrument Straight and related to programs on the DAC Universal load contra-angle Sterilization cycle type S handpieces and…
  • Page 34: Electromagnetic Compatibility

    Medical devices connected to the DAC Universal must comply with IEC 60601-1. Non-medical devices connected to the DAC Universal must comply with IEC 60601-1-1. The DAC Universal does not cause any harmful interference; it can be impaired by electromagnetic interference emitted by other devices, which can lead to malfunctions.

  • Page 35: Interference Immunity

    Sirona Dental A/S 3 Product description Operating Instructions DAC Universal 3.6 Electromagnetic compatibility 3.6.2 Interference immunity The UNIT is intended for operation in the electromagnetic environment specified below. The customer or user of the UNIT should make sure that it is used in such an environment.

  • Page 36
    3 Product description Sirona Dental A/S 3.6 Electromagnetic compatibility Operating Instructions DAC Universal Interference immunity IEC 60601-1-2 test Compliance level Electromagnetic environment — tests level guidelines Conducted RF d= [1.2] √P interference 150 kHz to 80 MHz IEC 61000-4-6 Radiated RF interference 3 V/m d= [1.2] √P…
  • Page 37: Working Clearances

    Sirona Dental A/S 3 Product description Operating Instructions DAC Universal 3.6 Electromagnetic compatibility 3.6.3 Working clearances Recommended working clearances The UNIT is intended for operation in an electromagnetic environment, between portable and mobile RF where radiated RF interference is checked. The customer or the user of…

  • Page 38: Safety Devices

    Watchdog relay When the microprocessor in the DAC Universal does not send a signal to the two watchdog relays, they switch off the power supply to the pumps, valves, motor, and steam generator.

  • Page 39: Installation

    Risk of crushing The recommended minimum height for installing the DAC Universal is 70 cm, in order to prevent injuries when removing the lid. If the DAC Universal needs to be switched off completely, the power plug must be accessible. Ambient conditions…

  • Page 40: Power Supply

    Operating Instructions DAC Universal 4.2.1 Power supply The DAC Universal must be connected to a supply voltage of 90 VAC – 120 VAC or 190 VAC – 240 VAC, 50 – 60 Hz, 1.100 W, at a grounded socket. NOTICE…

  • Page 41: Permanent Water Connection

    4.2.4 Drain connection NOTICE Do not start a cycle if the DAC Universal is not connected to the drain tank or drainage system. The DAC Universal works only in a one-way water system, i.e. used water (waste water) is not reused.

  • Page 42: Connecting The Drain Tank

    IMPORTANT A drain tank other than the original one must not be used. If the DAC Universal is not connected directly to the drainage system, a drain tank with a condensation tube must be installed to collect waste water from the reprocessing process.

  • Page 43: Connecting The Siphon

    1. Make sure that the drain system is clear and no counter pressure can build up. 2. The DAC Universal can be connected directly to the drainage system. Using the lubrication concentrate The NitramOil #2 bottle is not yet mounted when the unit is delivered. The bottle must be mounted and oil pumped through the hoses before the first use of the DAC Universal.

  • Page 44: Selecting The Display Language

    Operating Instructions DAC Universal Selecting the display language We recommend that the main language in your country or the language most familiar to the users of the DAC Universal is selected as display language. There are 6 languages available: ● English (UK) ●…

  • Page 45: Connecting To A Documentation Device

    (e.g. SegoSoft or DIOS). Printer For the DAC Universal an optional thermal printer is available. This is used for documenting the reprocessing cycle. See «Printer [ → 105]» for the technical data and version of the printer.

  • Page 46: Validation

     Printer language is set. 14. Press the «Clear»-Key until the start menu reappears. Validation The DAC Universal has been type tested and works tested according to requirements from applicable standards. The standards also specify installation and performance tests upon installation and use.

  • Page 47: Operation

    5.1 Regular activities Operation Regular activities The DAC Universal was installed by a service technician according to the instructions in the «Installation [ → 39]» section. See below for an overview of the activities that must be conducted during the regular use of the DAC Universal.

  • Page 48: Checking The Reprocessing Result With An Indicator / Pcd Test Device

    Use a commercial chemical indicator (Class 6), e.g. gke Steri-Record C- S-P-6-SV2 for the prion program. IMPORTANT All chemical indicators used in the DAC Universal must be coated or non-paper chemical indicators. IMPORTANT Follow the manufacturer’s instructions for use when using a chemical indicator.

  • Page 49
    If the chemical indicator indicates that the test was not passed, do not use the instruments that were reprocessed up to the time of the last successful check. Do not continue to use the DAC Universal. Please contact your authorized dealer. 5.1.2.1.1 Using a chemical indicator with the Standard lid 1.
  • Page 50
    5 Operation Sirona Dental A/S 5.1 Regular activities Operating Instructions DAC Universal 2. Attach the indicator holder,Standard lid in the groove on the center post of the Standard lid. 3. Start a cycle. See «Reprocessing with the Standard lid [ → 64]» for the detailed procedure.
  • Page 51: Testing With A Pcd Process Challenge Device

    The DAC Universal is switched on and ready for a cycle. ✔ Up to 5 instruments are ready for reprocessing in the DAC Universal, see «Pre- and post-reprocessing of instruments [ → 47]». 1. Remove the Standard lid from the holder on the DAC Universal.

  • Page 52
    5 Operation Sirona Dental A/S 5.1 Regular activities Operating Instructions DAC Universal 4. Fold the chemical indicator into a V-shape. 5. Put the chemical indicator into the PCD test device. 6. Twist the PCD test device clockwise to close. IMPORTANT Make sure the PCD test device is completely closed.
  • Page 53
    Sirona Dental A/S 5 Operation Operating Instructions DAC Universal 5.1 Regular activities 11. Remove the PCD test device from the adapter by pressing the small button at the bottom of the adapter while pulling on the PCD test device. CAUTION Risk of burns! The PCD test device may be hot.
  • Page 54: Adding/Replacing The Lubrication Concentrate

    Use only NitramOil #2 for the DAC Universal. IMPORTANT In case of fire: There is maximum 200 ml / 0.043 imperial gallons of oil in the DAC Universal. There is no danger that dangerous or harmful gases can develop in case of fire. 66 15 764 D3472…

  • Page 55
    (next to the water tank). 5. Turn the oil bottle carefully clockwise until you meet resistance. 6. Close the lid of the water tank.  The oil bottle is now inserted in the DAC Universal. 66 15 764 D3472 D3472.201.21.01.02 05.2017…
  • Page 56: Pumping Lubrication Concentrate Through The System

     The oil bottle has been replaced. Tip: You can also remove a partially used oil bottle that is not yet empty from the DAC Universal and insert it again at a later time. 5.1.4 Pumping lubrication concentrate through the system When a new bottle of lubrication concentrate has been inserted, you must check whether oil is pumped into the chamber.

  • Page 57
    Sirona Dental A/S 5 Operation Operating Instructions DAC Universal 5.1 Regular activities 1. Open the lid of the water tank. 2. Fill demineralized/distilled water (water conductivity < 3.0 µS/cm) through the sieve into the water tank (min. 1.1 liter, max. 2 liter).
  • Page 58: Using The Menu And Display Information

    The cycle is ended. The lid opens slightly. «Press C to unload» Indicates that the procedure is finished. Press the «Clear»-Key to open the DAC Universal completely so the instruments can be removed. «Service needed» (simultaneously 2 seconds of acoustic alarm is given) Indicates that 2 years or 3000 cycles have passed since the first use or last servicing.

  • Page 59: Menu System

    Sirona Dental A/S 5 Operation Operating Instructions DAC Universal 5.2 Using the menu and display information System status Specifications «Sterilization error» «Disinfection Sterilization/Disinfection has failed. This is displayed simultaneously with an error» acoustic alarm. The list of all display messages is provided in «Errors / Display messages [ →…

  • Page 60: Menu Overview

    5 Operation Sirona Dental A/S 5.2 Using the menu and display information Operating Instructions DAC Universal 5.2.2.2 Menu overview Menu level 1 Menu level 2 Menu level 3 Result / Activity / Note «Program» «Ready 134 °C» Program: 134 °C / 273 °F, Sterilization time 3 min.

  • Page 61
    Setting the minutes. «Settings» «Automatic «ON » Automatic filling of the water tank by a water conditioning water» system connected to the DAC Universal. «OFF» Manual filling of the water tank. «Settings» «Oil Settings» 1-3 oil settings Oil volume for lubrication of instruments.
  • Page 62: Default Settings

    5 Operation Sirona Dental A/S 5.2 Using the menu and display information Operating Instructions DAC Universal 5.2.3 Default settings The default settings are predetermined by the manufacturer: Program 134 °C / 273 °F, Sterilization time 3 min. Printer: Language: Oil expel: Oil settings:…

  • Page 63: Drain The Chamber

    DAC Universal is connected to a power socket, unless the DAC Universal was put into standby mode before having been disconnected from the power socket. In that case, the DAC Universal will restart in standby mode and must be started by pressing the «Start»-Key. IMPORTANT It is not possible to start a reprocessing program until the steam generator is properly heated.

  • Page 64: Reprocessing With The Standard Lid

    The instruments must not be wrapped when using the Standard lid, as cleaning, lubrication, and sterilization would then fail. IMPORTANT Do not start the DAC Universal without a lid inserted into the holder. ✔ All settings of the DAC Universal were preset according to «Regular activities [ →…

  • Page 65
    5 Operation Operating Instructions DAC Universal 5.3 Reprocessing with the DAC Universal 5. Insert the Standard lid in the holder of the DAC Universal in such a way that the instruments are pointing down. NOTICE Make sure that the lid is inserted all the way into the holder so that the instrument heads do not touch the wall of the chamber.
  • Page 66: D3472

    The metal part of the lid is very hot. At the end of the cycle, touch only the plastic part of the lid. 11. Remove the Standard lid from the holder on the DAC Universal at the end of the cycle to allow the instruments to cool down faster.

  • Page 67
    Sirona Dental A/S 5 Operation Operating Instructions DAC Universal 5.3 Reprocessing with the DAC Universal 16. Check whether the dark status bar of the chemical indicator is in the «ACCEPT» range.  If everything is correct, reprocessing was successful. 17. Document the cycle according to the work instructions applicable in your practice.
  • Page 68: Reprocessing With The Basket Lid

    The instruments must not be wrapped when using the Basket lid, as cleaning and sterilization would then fail. IMPORTANT Do not start the DAC Universal without a lid inserted into the holder. ✔ All settings of the DAC Universal were preset according to «Regular activities [ →…

  • Page 69
    Sirona Dental A/S 5 Operation Operating Instructions DAC Universal 5.3 Reprocessing with the DAC Universal 3. Open the snap lock on the basket and take the small lid off the basket. 4. Put maximum 5 solid instruments into the basket, if possible with the sharp tips pointing downwards.
  • Page 70
    Place a NitraClean tablet in the empty chamber. See «NitraClean cleaning tablet [ → 104]» for more information on using a NitraClean tablet. 8. Insert the Basket lid in the holder of the DAC Universal in such a way that the basket points down. NOTICE Make sure that the lid is inserted all the way into the holder so that the basket does not touch the wall of the chamber.
  • Page 71
    If you do not select a program, the last used program will be used. If you want to change the program, proceed as follows: ➢ When the Basket lid is inserted in the holder of the DAC Universal, press the «Menu/Enter»-Key twice and use the arrow keys to select the appropriate program.
  • Page 72
    The metal part of the lid is very hot. At the end of the cycle, touch only the plastic part of the lid. 15. Remove the Basket lid from the holder on the DAC Universal at the end of the cycle to allow the instruments to cool down faster.
  • Page 73: Reprocessing With The Flex Lid

    / PCD test device [ → 48]» for more information on using the chemical indicator. 5. Insert the Flex lid in the holder of the DAC Universal in such a way that the instruments are pointing down.

  • Page 74
    If you do not select a program, the last used program will be used. If you want to change the program, proceed as follows: ➢ When the Flex lid is inserted in the holder of the DAC Universal, press the «Menu/Enter»-Key twice and use the arrow keys to select the appropriate program.
  • Page 75
    The metal part of the lid is very hot. At the end of the cycle, touch only the plastic part of the lid. 12. Remove the Flex lid from the holder on the DAC Universal at the end of the cycle to allow the instruments to cool down faster.
  • Page 76
    5 Operation Sirona Dental A/S 5.3 Reprocessing with the DAC Universal Operating Instructions DAC Universal  After the end of the cycle, the DAC Universal automatically reactivates lubrication for the next cycle. 66 15 764 D3472 D3472.201.21.01.02 05.2017…
  • Page 77: Cleaning And Maintenance

    Sirona Dental A/S 6 Cleaning and maintenance Operating Instructions DAC Universal Cleaning and maintenance After every 50 cycles, the DAC Universal will give a 2 second acoustic «Remember: alarm and the display will show the following information: » User Maintenance» » — Check & Clean. This message reminds the user that the weekly cleaning / maintenance according to the Check &…

  • Page 78: Checking Oil Supply To The Chamber

    Checking water supply to the chamber Frequency: weekly or every 50 cycles (whichever comes first). In order to make sure that the DAC Universal is able to bring water to the chamber correctly, this operation needs to be tested. Pump water into the chamber ✔…

  • Page 79: Checking Water Supply To The Lids

    Checking water supply to the lids Frequency: weekly or every 50 cycles (whichever comes first). In order to make sure that the DAC Universal is able to bring water to the chamber and internal channels of the Standard and Flex lids correctly, this operation needs to be tested.

  • Page 80: Checking Oil Supply To The Standard Lid

    Checking oil supply to the Standard lid Frequency: weekly or every 50 cycles (whichever comes first). In order to make sure that the DAC Universal is able to lubricate the instruments correctly, the flow through the internal oil channels of the Standard lid needs to be tested.

  • Page 81: Drain The Water Tank

    (if using a drain tank). ✔ The Standard lid is at hand and cooled down. 1. Insert an empty Standard lid into the holder of the DAC Universal. 2. Press the «Menu/Enter»-Key to go to menu level 1. «Manual»…

  • Page 82: Cleaning The Water Tank

    Depending on the amount of water in the water tank, this will take approx. 10-40 min. IMPORTANT If you have installed the DAC Universal with a permanent water connection, the automatic water function must be turned on again when you have drained the tank, see «Cleaning the water tank [ → 82]».

  • Page 83
    Operating Instructions DAC Universal 6.6 Cleaning the water tank 4. Lift the water tank straight out of the DAC Universal. 5. Clean the water tank with warm water (maximum 50 ºC / 122 ºF). 6. Clean the sieve with warm water (maximum 50 ºC / 122 ºF).
  • Page 84
    14. For manual water supply: Fill demineralized or distilled water (water conductivity < 3.0 µS/cm) into the water tank. 15. Reconnect the DAC Universal to the power supply.  The DAC Universal is now switched on and ready for operation again. Turn on «automatic water» IMPORTANT Only turn on automatic water if the DAC Universal is connected to a permanent water conditioning system.
  • Page 85: Cleaning The Drain Filter

     When a cycle is started, error 81, «low water level», will appear on the display until the water level is sufficient.  The DAC Universal will then be ready for use. IMPORTANT If there is no permanent water connection, fill the water tank with demineralized or distilled water (water conductivity <…

  • Page 86: Removing And Cleaning The Drain Tank

    If using automatic water: Daily or every 50 cycles (whichever comes first) or when it is full. NOTICE Do not start a cycle if the DAC Universal is not connected to the drain tank or drainage system. ✔ The DAC Universal is switched off and cooled down.

  • Page 87
    Sirona Dental A/S 6 Cleaning and maintenance Operating Instructions DAC Universal 6.8 Removing and cleaning the drain tank 1. Unscrew the lid of the drain tank. 2. Remove the condensation tube. 3. Empty the drain tank. Dispose of the waste water according to the national guidelines and the applicable regulations for waste disposal.
  • Page 88: Checking The Adapters And The Respective O-Rings

    This prevent the growth of bacteria in the drain tank. 8. Replace the condensation tube and screw the lid on firmly. 9. Place the drain tank no more than 3 m from the DAC Universal and at a lower height.

  • Page 89: Cleaning The Empty Chamber

    ✔ The DAC Universal is switched on and ready for operation. ✔ The Standard lid of the DAC Universal is removed and there is an unobstructed view into the chamber. ✔ There is no water or oil in the chamber.

  • Page 90
    20 seconds, you must press the «Start»-Key and choose the lid again.  The Standard lid of the DAC Universal closes as soon as the safety ring is pressed.  The DAC Universal automatically runs the cleaning and sterilization cycle.
  • Page 91: Cleaning The Housing Of The Unit

    The metal part of the lid is very hot. At the end of the cycle, touch only the plastic part of the lid. 10. Remove the Standard lid from the holder on the DAC Universal at the end of the cycle.

  • Page 92: Checking The Safety Ring

    Frequency: every 4 weeks or every 200 cycles (whichever comes first). ✔ The DAC Universal is switched on and ready for operation. 1. Insert the Standard lid in the holder of the DAC Universal. NOTICE No instruments must be on the Standard lid.

  • Page 93: Changing The Drain Filter

    6 Cleaning and maintenance Operating Instructions DAC Universal 6.14 Changing the drain filter CAUTION If the check does not proceed as described above, the DAC Universal must not continue to be used. Please contact your authorized dealer. 6.14 Changing the drain filter Frequency: every 4 weeks or every 200 cycles (whichever comes first).

  • Page 94: Changing The O-Rings On The Adapters

    Frequency: every 2nd year or every 3000 cycles (whichever comes first). When 2 years or 3000 cycles have passed, since the last service or the DAC Universal was put into use for the first time, the screen «Service needed» will be displayed, simultaneously with a 2 second acoustic alarm.

  • Page 95: Errors / Display Messages

    CAUTION Risk of burns! Hot steam may emerge from the chamber. ➢ Do not open the DAC Universal as long as the temperature is above 104 °C (219 °F). Refer to the display. IMPORTANT If the DAC Universal shows the same display message more than once, and this is not due to lack of consumables, please contact your authorized dealer.

  • Page 96: Display Messages And What They Mean

    Code Specifications Corrective action The heat capacity of the steam If the DAC Universal is in a room with a temperature < 16 °C / 61 °F, generator was not regained ensure that operating conditions are fulfilled. during the cycle.

  • Page 97
    110 °C / 230 °F. conductivity < 3.0 µS/cm) into the water tank. If the DAC Universal is in a room with a temperature < 16 °C / 61 °F, ensure that operating conditions are fulfilled. Press the «Clear»-Key to reset and then restart the DAC Universal.
  • Page 98
    120 °C / 248 °F conductivity < 3.0 µS/cm) into the water tank. was not reached within 2.5 If the DAC Universal is in a room with a temperature < 16 °C / 61 °F, minutes. ensure that operating conditions are fulfilled.
  • Page 99
    If the error occurs repeatedly, please contact your authorized dealer. «Clear»-Key was activated during the cycle. The steam generator is Switch the DAC Universal off immediately by pulling the power cord out of overheated. the socket. Please contact your authorized dealer.
  • Page 100
    Operating Instructions DAC Universal Code Specifications Corrective action Oil consumption during standby. Disconnect the DAC Universal from the power supply and wait 10 seconds. Then reconnect the DAC Universal to the power supply and restart. The DAC Universal failed to fill Important: This error can occur only if the DAC Universal is connected to the water tank within 4 minutes.
  • Page 101: Errors Without A Message

    Sterilization OFF safety ring! (A long Your DAC Universal technician has run a special program. acoustic alarm will be given Disconnect the plug of the DAC Universal and reconnect it to the power simultaneously). supply. WARNING: Otherwise no sterilization will take place! The DAC Universal does not open.

  • Page 102: Reset

    Reset If the DAC Universal was started without attaching the lid in the holder of the DAC Universal and if the holder pressed the safety ring down and can no longer be opened, please proceed as follows: ✔…

  • Page 103: Accessories

    Water filter (20µm) 6078575 Air filter (20µm) 6300623 Sterile filter 6051770 DAC Universal thermal printer 6123462 DAC Universal thermal printer paper, (5 rolls) 6051820 PCD test device 6051788 Indicator holder, Standard lid 6542489 Indicator holder, Flex lid 6078526 Drain tank with hose…

  • Page 104: Nitraclean Cleaning Tablet

    8 Accessories Sirona Dental A/S 8.1 NitraClean cleaning tablet Operating Instructions DAC Universal For further information about the adapters and the corresponding O-rings, please see the separate operating instructions for adapters. NitraClean cleaning tablet NitraClean tablets are used for: ● Cleaning solid instruments in the Basket lid ●…

  • Page 105: Printer

    Technical Data ● Serial interface: RS232 Output information Standard lid and Basket lid: The following information is given by the printer: ● Serial number of the DAC Universal ● Program ● Cycle number ● Date (DD-MM-YYYY) and time (HH:MM:SS) (Start) ●…

  • Page 106
    If any of the phases failed, see «Errors / Display messages [ → 95]». Output information Flex lid: The following information is given by the printer: ● Serial number of the DAC Universal ● Program ● Cycle number ● Date (DD-MM-YYYY) and time (HH:MM:SS) (Start) ●…
  • Page 107: Consumables, Spare Parts, And Tools

    Sirona Dental A/S 9 Consumables, spare parts, and tools Operating Instructions DAC Universal Consumables, spare parts, and tools Consumables Article designation 6259118 NitramOil #2, lubrication concentrate (blue), (6 bottles) ** 6053842 NitraClean cleaning tablets — for cleaning the chamber 6078583 Drain filter for filter housing (pack of 6)

  • Page 108: Disposal

    10 Disposal Sirona Dental A/S Operating Instructions DAC Universal Disposal In accordance with Directive 2012/19/EU and national disposal regulations regarding old electrical and electronic devices, please be advised that such items must be disposed of in a special way within the European Union (EU).

  • Page 110
    tÉ=êÉëÉêîÉ=íÜÉ=êáÖÜí=íç=ã~âÉ=~åó=~äíÉê~íáçåë=ïÜáÅÜ=ã~ó=ÄÉ=êÉèìáêÉÇ=ÇìÉ=íç=íÉÅÜåáÅ~ä=áãéêçîÉãÉåíëK «=páêçå~=aÉåí~ä==^Lp=OMNT péê~ÅÜÉW ÉåÖäáëÅÜ mêáåíÉÇ=áå=aÉåã~êâ aPQTOKOMNKONKMNKMO MRKOMNT ûKJkêKW= ahMRQM páêçå~=aÉåí~ä=^Lp SS=NR=TSQ=aPQTO oÜç=NM lêÇÉê=kç UPUO=eáååÉêìé bk|RNMUQSJORi|p aÉåã~êâ 0088 ïïïKëáêçå~KÅçã…

(Ocr-Read Summary of Contents of some pages of the Sirona DAC UNIVERSAL Document (Main Content), UPD: 27 August 2023)

  • 151, 60 53 362 D3472 D3472.201.10.20.02 07.2016 151 Sirona Dental A/S 7Display messages DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL 7.2Errors without a message båÖäáëÜ 7.2 Errors without a message Description Corrective action Leakage. Wipe away any water that may have been spilled after filling. If the unit continues to leak, please contact your authorized dealer. There is water in the chamber aft…

  • 195, Sirona DAC UNIVERSAL 60 53 362 D3472 D3472.201.10.20.02 07.2016 195 Sirona Dental A/S 9Cleaning and maintenance DAC UNIVERSAL STANDARD Operating Instructions DAC UNIVERSAL 9.4Checking oil supply to the STANDARD lid of the DAC UNIVERSAL STANDARD båÖäáëÜ 3. Assure the lid is attached vertically on the bottom adapter, as shown. 4. Press the «Rinse/Flush»-Key and hold it down for more than 1 second.  Water is pumped through the internal water h…

  • 13, 60 53 362 D3472 D3472.201.10.20.02 07.2016 13 Sirona Dental A/S 1General Information Operating Instructions DAC UNIVERSAL 1.8Intended use båÖäáëÜ 1.8 Intended use The DAC UNIVERSAL is intended to clean, lubricate and steam sterilize dental handpieces. Dental handpieces is defined as rotating instruments split up into the categories straight handpieces, contra angles and turbines. The DAC UNIVERSAL is in…

  • 202, 60 53 362 D3472 202 D3472.201.10.20.02 07.2016 9Cleaning and maintenance DAC UNIVERSAL STANDARD Sirona Dental A/S 9.8Remove and clean the drain tank Operating Instructions DAC UNIVERSAL 9.8 Remove and clean the drain tank Frequency: If using manual water: Weekly or every 50 cycles (whichever comes first) or when it is full. If using automatic water: Daily or every 50 cycles (whichever comes first) or when it is full. ✔ The DAC UNIVERSAL is switched off and cooled dow…

  • 157, 60 53 362 D3472 D3472.201.10.20.02 07.2016 157 Sirona Dental A/S 8Operation DAC UNIVERSAL STANDARD Operating Instructions DAC UNIVERSAL 8.1Regular activities båÖäáëÜ 2. Attach the STANDARD indicator holder to one of the three wire brackets of the WRAPPED lid so the brackets of the indicator holder point to the center of the lid. 3. Press the STANDARD indicator holder toward the lid as far as possible. 4. Attach a wrapped instrument between the wires at each of the…

  • 124, Sirona DAC UNIVERSAL 60 53 362 D3472 124 D3472.201.10.20.02 07.2016 6Cleaning and maintanance DAC UNIVERSAL ADVANCED Sirona Dental A/S 6.4Checking oil supply to the STANDARD lid of the DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL  Water is pumped through the internal water hoses and channels of the lid. 6. When water is visible from underneath the lid’s top plate; observe that water is flowing freely from the 6 different adapters. 7. Repress the «Down“-Key. 8. Remove the black e…

  • 161, 60 53 362 D3472 D3472.201.10.20.02 07.2016 161 Sirona Dental A/S 8Operation DAC UNIVERSAL STANDARD Operating Instructions DAC UNIVERSAL 8.1Regular activities båÖäáëÜ Place safety data sheet_Sirona Inserting the oil bottle ✔ The lid of the DAC UNIVERSAL water tank is open. 1. Remove the red transport plug on the oil dispenser. Please keep this plug (it can be placed on the bottle after it is screwed i…

  • 212, 60 53 362 D3472 212 D3472.201.10.20.02 07.2016 10Display messages DAC UNIVERSAL STANDARD Sirona Dental A/S 10.1Display messages and what they mean DAC UNIVERSAL STANDARD Operating Instructions DAC UNIVERSAL 10.1 Display messages and what they mean DAC UNIVERSAL STANDARD The display messages have the following meanings: Code no. Description Corrective action 60 The chamber temperature and the theoretical steam temperature devi…

  • 2, 60 53 362 D3472 2 D3472.201.10.20.02 07.2016 Dear Customer, Sirona Dental A/S Operating Instructions DAC UNIVERSAL Dear Customer, Thank you for purchasing a DAC UNIVERSAL unit. DAC UNIVERSAL cleans, oils and sterilizes / disinfects 6 instruments automatically. Wrapped and solid instruments can also be sterilized in next to no time and are therefore back in action quickly. DAC UNIVERSAL saves you time-consuming manual reprocessing and reliably protects you and your …

  • 5, 60 53 362 D3472 D3472.201.10.20.02 07.2016 5 Sirona Dental A/S Table of Contents Operating Instructions DAC UNIVERSAL båÖäáëÜ 5.1.2.1 Testing with a chemical indicator……………………………….. 60 5.1.2.2 Testing with a PCD process challenge device …………….. 64 5.1.2.3 Testing with a protein residue test……………………………… 66 5.1.3 Adding/replacing the lubrication concentrate ………………………

  • 231, 60 53 362 D3472 D3472.201.10.20.02 07.2016 231 Sirona Dental A/S 11Accessories Operating Instructions DAC UNIVERSAL 11.1Adapters for the STANDARD lid båÖäáëÜ 11.1.1.12 Adapter for Borden 2/3-hole fixed connection Dealing with adapters_Sirona_Nitram Follow the general safety information in «General placement of instruments DAC UNIVERSAL ADVANCED [ → 99]» or «General placement of instruments DAC UNIVERSAL STANDARD [ → 172…

  • 246, 60 53 362 D3472 246 D3472.201.10.20.02 07.2016 12Consumables, spare parts, and tools Sirona Dental A/S Operating Instructions DAC UNIVERSAL * Internal O-rings on adapters may be replaced only by qualified service technicians * Internal O-rings on adapters may be replaced only by qualified service technicians * Internal O-rings on adapters may be replaced only by qualified service technicians Consumables ** SN 104,000-199,999 & 220,000-399,999 Tools for …

  • 237, Sirona DAC UNIVERSAL 60 53 362 D3472 D3472.201.10.20.02 07.2016 237 Sirona Dental A/S 11Accessories Operating Instructions DAC UNIVERSAL 11.1Adapters for the STANDARD lid båÖäáëÜ 11.1.1.18 Adapter for EMS ultrasonic scaler with light Dealing with adapters_Sirona_Nitram Follow the general safety information in «General placement of instruments DAC UNIVERSAL ADVANCED [ → 99]» or «General placement of instruments DAC UNIVERSAL STANDARD…

  • 82, 60 53 362 D3472 82 D3472.201.10.20.02 07.2016 5Operation DAC UNIVERSAL ADVANCED Sirona Dental A/S 5.2Using the menu and display information DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL 70% 80% 90% 100% Automatic water Automatic water ON Automatic water OFF History Cycles Cycle history Service Next service Errorbuffer Errors 1-3 Errors 4-6 Errors 7-9 Level 1 Level 2 Level 3 Level 4 Level 5 Description

  • 69, Sirona DAC UNIVERSAL 60 53 362 D3472 D3472.201.10.20.02 07.2016 69 Sirona Dental A/S 5Operation DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL 5.1Regular activities båÖäáëÜ 5.1.5 Filling the water tank If no permanent water connection is selected, the DAC UNIVERSAL water tank must be filled with water manually. 1 filled water tank corresponds to approx. 2 complete cycles of the standard program. Capacity of the water tan…

  • 109, 60 53 362 D3472 D3472.201.10.20.02 07.2016 109 Sirona Dental A/S 5Operation DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL 5.3Reprocessing with the DAC UNIVERSAL ADVANCED båÖäáëÜ must be conducted in the DAC UNIVERSAL or a class B sterilizer (e.g. DAC PROFESSIONAL). For sterilizing wrapped instruments in DAC UNIVERSAL, see «Reprocessing with the WRAPPED lid DAC UNIVERSAL ADVANCED [ → 110]».

  • 28, 60 53 362 D3472 28 D3472.201.10.20.02 07.2016 3Product description Sirona Dental A/S 3.2Controls and functional elements of the DAC UNIVERSAL Operating Instructions DAC UNIVERSAL 3.2.2 Lid The following lids can be used for reprocessing in the DAC UNIVERSAL: STANDARD lid_DAC U. ADVANCED ● STANDARD lid for reprocessing unwrapped rotating handpieces. See «Functions with the STANDARD lid [ → 20]» for functions and «Reprocessing with the ST…

  • 223, 60 53 362 D3472 D3472.201.10.20.02 07.2016 223 Sirona Dental A/S 11Accessories Operating Instructions DAC UNIVERSAL 11.1Adapters for the STANDARD lid båÖäáëÜ 11.1.1.4 Adapter for W&H Roto Quick Dealing with adapters_Sirona_Nitram Follow the general safety information in «General placement of instruments DAC UNIVERSAL ADVANCED [ → 99]» or «General placement of instruments DAC UNIVERSAL STANDA…


3.0

Rated 3 out of 5

3 out of 5 stars (based on 1 review)

Your overall rating

DENTSPLY SIRONA DAC UNIVERSAL (01) PDF MANUAL

Click here to download DENTSPLY SIRONA DAC UNIVERSAL (01) PDF MANUAL


DENTSPLY SIRONA DAC UNIVERSAL (01) PDF MANUAL

FREE ENGLISH PDF

OPERATING INSTRUCTIONS

USER GUIDE – USER MANUAL

OWNER GUIDE – OWNER MANUAL

REFERENCE GUIDE – REFERENCE MANUAL

INSTRUCTION GUIDE – INSTRUCTION MANUAL


Your overall rating

  • Twitter
  • YouTube
  • Reddit
  • Facebook
  • Instagram
  • Pinterest

DENTSPLY SIRONA DAC UNIVERSAL (01) PDF MANUAL

DENTSPLY SIRONA DAC UNIVERSAL (01) PDF MANUAL

[Page 1] Sirona DAC UNIVERSAL

kÉï=~ë=çÑW= léÉê~íáåÖ=fåëíêìÅíáçåë MRKOMNT a^`=råáîÉêë~ä qÉñí=aáëéä~ó=J=sÉêëáçå=OR båÖäáëÜ = Cover page

[Page 2] Sirona DAC UNIVERSAL

66 15 764 D3472 2 D3472.201.21.01.02 05.2017 Dear Customer, Sirona Dental A/S Operating Instructions DAC Universal Dear Customer, Thank you for purchasing a DAC Universal. The DAC Universal cleans, lubricates and disinfects / sterilizes 6 instrumen…

[Page 3] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 3 Sirona Dental A/S Table of Contents Operating Instructions DAC Universal båÖäáëÜ Table of Contents 1 General Information……………………………………………………………………….

[Page 4] Sirona DAC UNIVERSAL

66 15 764 D3472 4 D3472.201.21.01.02 05.2017 Table of Contents Sirona Dental A/S Operating Instructions DAC Universal 3.5 Presentation of the sterilization process ………………………………………………. 29 3.6 Electromagnetic compat…

[Page 5] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 5 Sirona Dental A/S Table of Contents Operating Instructions DAC Universal båÖäáëÜ 5.2.5 Drain the chamber………………………………………………………………… 63 5.3 Reprocessing…

[Page 6] Sirona DAC UNIVERSAL

66 15 764 D3472 6 D3472.201.21.01.02 05.2017 1General Information Sirona Dental A/S 1.1Contact information Operating Instructions DAC Universal 1 General Information 1.1 Contact information Customer Service Center For technical questions you c…

[Page 7] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 7 Sirona Dental A/S 1General Information Operating Instructions DAC Universal 1.5Structure of the document båÖäáëÜ 1.5 Structure of the document 1.5.1 Identification of the danger levels To preve…

[Page 8] Sirona DAC UNIVERSAL

66 15 764 D3472 8 D3472.201.21.01.02 05.2017 1General Information Sirona Dental A/S 1.6Warranty and liability Operating Instructions DAC Universal 1.6 Warranty and liability The manufacturer guarantees that the product delivered to you (DAC U…

[Page 9] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 9 Sirona Dental A/S 1General Information Operating Instructions DAC Universal 1.7Intended use båÖäáëÜ 1.7 Intended use The DAC Universal is intended to clean, lubricate and steam sterilize denta…

[Page 10] Sirona DAC UNIVERSAL

66 15 764 D3472 10 D3472.201.21.01.02 05.2017 1General Information Sirona Dental A/S 1.9Scope of delivery Operating Instructions DAC Universal 1.9 Scope of delivery The following components are included in the scope of delivery of the DAC Uni…

[Page 11] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 11 Sirona Dental A/S 2Safety instructions Operating Instructions DAC Universal 2.1Electrostatic discharge båÖäáëÜ 2 Safety instructions 2.1 Electrostatic discharge Protective measures Electrostat…

[Page 12] Sirona DAC UNIVERSAL

66 15 764 D3472 12 D3472.201.21.01.02 05.2017 2Safety instructions Sirona Dental A/S 2.1Electrostatic discharge Operating Instructions DAC Universal An electrostatic discharge is: ● perceptible at 3,000 V or higher ● audible at 5,000 V or …

[Page 13] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 13 Sirona Dental A/S 2Safety instructions Operating Instructions DAC Universal 2.2Caution symbols on the device båÖäáëÜ 2.2 Caution symbols on the device Meaning: Warning of danger This symbol is…

[Page 14] Sirona DAC UNIVERSAL

66 15 764 D3472 14 D3472.201.21.01.02 05.2017 2Safety instructions Sirona Dental A/S 2.3Flammable materials Operating Instructions DAC Universal 2.3 Flammable materials 2.4 Reducing the risk of burns CAUTION Risk of explosion Never use the DAC…

[Page 15] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 15 Sirona Dental A/S 2Safety instructions Operating Instructions DAC Universal 2.5Instruction of practice personnel båÖäáëÜ 2.5 Instruction of practice personnel The personnel responsible for ope…

[Page 16] Sirona DAC UNIVERSAL

66 15 764 D3472 16 D3472.201.21.01.02 05.2017 3Product description Sirona Dental A/S 3.1Functions Operating Instructions DAC Universal 3 Product description 3.1 Functions The DAC Universal reduces the risk of cross-contamination and ensures t…

[Page 17] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 17 Sirona Dental A/S 3Product description Operating Instructions DAC Universal 3.1Functions båÖäáëÜ Internal cleaning Internal cleaning ensures that the channels of the instruments are rinsed th…

[Page 18] Sirona DAC UNIVERSAL

66 15 764 D3472 18 D3472.201.21.01.02 05.2017 3Product description Sirona Dental A/S 3.1Functions Operating Instructions DAC Universal Sterilization After external cleaning, the chamber is heated to 134 °C (273 °F) and the instruments are s…

[Page 19] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 19 Sirona Dental A/S 3Product description Operating Instructions DAC Universal 3.1Functions båÖäáëÜ See «NitraClean cleaning tablet [ → 104]» for more information on NitraClean clean…

[Page 20] Sirona DAC UNIVERSAL

66 15 764 D3472 20 D3472.201.21.01.02 05.2017 3Product description Sirona Dental A/S 3.1Functions Operating Instructions DAC Universal Internal cleaning Internal cleaning ensures that the channels of the instruments are rinsed thoroughly. For…

[Page 21] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 21 Sirona Dental A/S 3Product description Operating Instructions DAC Universal 3.2Controls and functional elements båÖäáëÜ Find enclosed the characteristic values of the Flex lid. To select the p…

[Page 22] Sirona DAC UNIVERSAL

66 15 764 D3472 22 D3472.201.21.01.02 05.2017 3Product description Sirona Dental A/S 3.2Controls and functional elements Operating Instructions DAC Universal 3.2.2 Lid The following lids can be used for reprocessing in the DAC Universal: ● S…

[Page 23] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 23 Sirona Dental A/S 3Product description Operating Instructions DAC Universal 3.2Controls and functional elements båÖäáëÜ 3.2.3 Display and key pad A I H G F E D C B Display/button Specificati…

[Page 24] Sirona DAC UNIVERSAL

66 15 764 D3472 24 D3472.201.21.01.02 05.2017 3Product description Sirona Dental A/S 3.2Controls and functional elements Operating Instructions DAC Universal 3.2.4 Back of unit See also «Installation of connections [ → 39]». 3.2.5 …

[Page 25] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 25 Sirona Dental A/S 3Product description Operating Instructions DAC Universal 3.2Controls and functional elements båÖäáëÜ 3.2.6 Water tank with sieve The water tank is used for manually filling …

[Page 26] Sirona DAC UNIVERSAL

66 15 764 D3472 26 D3472.201.21.01.02 05.2017 3Product description Sirona Dental A/S 3.3Programs and program selection Operating Instructions DAC Universal 3.3 Programs and program selection The DAC Universal has 2 different programs that can …

[Page 27] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 27 Sirona Dental A/S 3Product description Operating Instructions DAC Universal 3.4Technical data båÖäáëÜ 3.4 Technical data Manufacturer: Sirona Dental A/S Model: DAC Universal Dimensions (H/W/D)…

[Page 28] Sirona DAC UNIVERSAL

66 15 764 D3472 28 D3472.201.21.01.02 05.2017 3Product description Sirona Dental A/S 3.4Technical data Operating Instructions DAC Universal The DAC Universal complies with the requirements of the RoHS Directive 2011/65/EU. Symbols on the DAC …

[Page 29] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 29 Sirona Dental A/S 3Product description Operating Instructions DAC Universal 3.5Presentation of the sterilization process båÖäáëÜ 3.5 Presentation of the sterilization process The following gra…

[Page 30] Sirona DAC UNIVERSAL

66 15 764 D3472 30 D3472.201.21.01.02 05.2017 3Product description Sirona Dental A/S 3.5Presentation of the sterilization process Operating Instructions DAC Universal Prion program The following graph shows temperature (°C) and pressure (bar)…

[Page 31] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 31 Sirona Dental A/S 3Product description Operating Instructions DAC Universal 3.5Presentation of the sterilization process båÖäáëÜ Basket lid The following graph shows temperature (°C) and pres…

[Page 32] Sirona DAC UNIVERSAL

66 15 764 D3472 32 D3472.201.21.01.02 05.2017 3Product description Sirona Dental A/S 3.5Presentation of the sterilization process Operating Instructions DAC Universal Flex lid The following graph shows temperature (°C) and pressure (bar) duri…

[Page 33] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 33 Sirona Dental A/S 3Product description Operating Instructions DAC Universal 3.5Presentation of the sterilization process båÖäáëÜ The table below gives an overview of what is performed during w…

[Page 34] Sirona DAC UNIVERSAL

66 15 764 D3472 34 D3472.201.21.01.02 05.2017 3Product description Sirona Dental A/S 3.6Electromagnetic compatibility Operating Instructions DAC Universal 3.6 Electromagnetic compatibility In order to prevent possible risks caused by electroma…

[Page 35] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 35 Sirona Dental A/S 3Product description Operating Instructions DAC Universal 3.6Electromagnetic compatibility båÖäáëÜ 3.6.2 Interference immunity The UNIT is intended for operation in the elect…

[Page 36] Sirona DAC UNIVERSAL

66 15 764 D3472 36 D3472.201.21.01.02 05.2017 3Product description Sirona Dental A/S 3.6Electromagnetic compatibility Operating Instructions DAC Universal 1. The higher frequency range applies at 80 MHz and 800 MHz. 2. The field strengths of f…

[Page 37] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 37 Sirona Dental A/S 3Product description Operating Instructions DAC Universal 3.6Electromagnetic compatibility båÖäáëÜ 3.6.3 Working clearances Recommended working clearances between portable a…

[Page 38] Sirona DAC UNIVERSAL

66 15 764 D3472 38 D3472.201.21.01.02 05.2017 3Product description Sirona Dental A/S 3.7Safety devices Operating Instructions DAC Universal 3.7 Safety devices Safety valve The safety valve was developed according to the EN 4126-1 standard and…

[Page 39] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 39 Sirona Dental A/S 4Installation Operating Instructions DAC Universal 4.1Requirements of the location / Ambient conditions båÖäáëÜ 4 Installation 4.1 Requirements of the location / Ambient cond…

[Page 40] Sirona DAC UNIVERSAL

66 15 764 D3472 40 D3472.201.21.01.02 05.2017 4Installation Sirona Dental A/S 4.2Installation of connections Operating Instructions DAC Universal 4.2.1 Power supply The DAC Universal must be connected to a supply voltage of 90 VAC – 120 VA…

[Page 41] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 41 Sirona Dental A/S 4Installation Operating Instructions DAC Universal 4.2Installation of connections båÖäáëÜ ● Permanent water connection. The DAC Universal is connected to a permanent water…

[Page 42] Sirona DAC UNIVERSAL

66 15 764 D3472 42 D3472.201.21.01.02 05.2017 4Installation Sirona Dental A/S 4.2Installation of connections Operating Instructions DAC Universal 4.2.4.1 Connecting the drain tank If the DAC Universal is not connected directly to the drainage …

[Page 43] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 43 Sirona Dental A/S 4Installation Operating Instructions DAC Universal 4.3Using the lubrication concentrate båÖäáëÜ 4.2.4.2 Connecting the siphon For connecting the drain connection with the sip…

[Page 44] Sirona DAC UNIVERSAL

66 15 764 D3472 44 D3472.201.21.01.02 05.2017 4Installation Sirona Dental A/S 4.5Selecting the display language Operating Instructions DAC Universal 4.5 Selecting the display language We recommend that the main language in your country or the …

[Page 45] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 45 Sirona Dental A/S 4Installation Operating Instructions DAC Universal 4.6Connecting to a documentation device båÖäáëÜ 4.6 Connecting to a documentation device The cycle can be documented with a…

[Page 46] Sirona DAC UNIVERSAL

66 15 764 D3472 46 D3472.201.21.01.02 05.2017 4Installation Sirona Dental A/S 4.7Validation Operating Instructions DAC Universal 5. Press the «Up» or «Down»-Key until «Printer» is selected. 6. Press the «Menu…

[Page 47] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 47 Sirona Dental A/S 5Operation Operating Instructions DAC Universal 5.1Regular activities båÖäáëÜ 5 Operation 5.1 Regular activities The DAC Universal was installed by a service technician acc…

[Page 48] Sirona DAC UNIVERSAL

66 15 764 D3472 48 D3472.201.21.01.02 05.2017 5Operation Sirona Dental A/S 5.1Regular activities Operating Instructions DAC Universal 5.1.2 Checking the reprocessing result with an indicator / PCD test device Installation / major repairs: Be…

[Page 49] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 49 Sirona Dental A/S 5Operation Operating Instructions DAC Universal 5.1Regular activities båÖäáëÜ For the Standard lid, the «indicator holder, Standard lid» is used, see «Using …

[Page 50] Sirona DAC UNIVERSAL

66 15 764 D3472 50 D3472.201.21.01.02 05.2017 5Operation Sirona Dental A/S 5.1Regular activities Operating Instructions DAC Universal 2. Attach the indicator holder,Standard lid in the groove on the center post of the Standard lid. 3. Start …

[Page 51] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 51 Sirona Dental A/S 5Operation Operating Instructions DAC Universal 5.1Regular activities båÖäáëÜ 2. Attach the indicator holder,Flex lid in the groove on the center post of the Flex lid. 3. S…

[Page 52] Sirona DAC UNIVERSAL

66 15 764 D3472 52 D3472.201.21.01.02 05.2017 5Operation Sirona Dental A/S 5.1Regular activities Operating Instructions DAC Universal 4. Fold the chemical indicator into a V-shape. 5. Put the chemical indicator into the PCD test device. 6. T…

[Page 53] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 53 Sirona Dental A/S 5Operation Operating Instructions DAC Universal 5.1Regular activities båÖäáëÜ 11. Remove the PCD test device from the adapter by pressing the small button at the bottom of …

[Page 54] Sirona DAC UNIVERSAL

66 15 764 D3472 54 D3472.201.21.01.02 05.2017 5Operation Sirona Dental A/S 5.1Regular activities Operating Instructions DAC Universal 5.1.3 Adding/replacing the lubrication concentrate The NitramOil #2 bottle is not yet mounted when the unit …

[Page 55] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 55 Sirona Dental A/S 5Operation Operating Instructions DAC Universal 5.1Regular activities båÖäáëÜ Inserting the oil bottle ✔ The lid of the DAC Universal water tank is open. 1. Remove the red…

[Page 56] Sirona DAC UNIVERSAL

66 15 764 D3472 56 D3472.201.21.01.02 05.2017 5Operation Sirona Dental A/S 5.1Regular activities Operating Instructions DAC Universal Replacement of the oil bottle ✔ The lid of the DAC Universal water tank is open. 1. Unscrew the empty oil …

[Page 57] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 57 Sirona Dental A/S 5Operation Operating Instructions DAC Universal 5.1Regular activities båÖäáëÜ 1. Open the lid of the water tank. 2. Fill demineralized/distilled water (water conductivity &l…

[Page 58] Sirona DAC UNIVERSAL

66 15 764 D3472 58 D3472.201.21.01.02 05.2017 5Operation Sirona Dental A/S 5.2Using the menu and display information Operating Instructions DAC Universal 5.2 Using the menu and display information 5.2.1 Display information The display indicat…

[Page 59] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 59 Sirona Dental A/S 5Operation Operating Instructions DAC Universal 5.2Using the menu and display information båÖäáëÜ 5.2.2 Menu system. 5.2.2.1 Navigation in the menu structure 1. Press the &q…

[Page 60] Sirona DAC UNIVERSAL

66 15 764 D3472 60 D3472.201.21.01.02 05.2017 5Operation Sirona Dental A/S 5.2Using the menu and display information Operating Instructions DAC Universal 5.2.2.2 Menu overview Menu level 1 Menu level 2 Menu level 3 Result / Activity / Note &q…

[Page 61] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 61 Sirona Dental A/S 5Operation Operating Instructions DAC Universal 5.2Using the menu and display information båÖäáëÜ «Settings» «Set time/date» «Set year» (00…

[Page 62] Sirona DAC UNIVERSAL

66 15 764 D3472 62 D3472.201.21.01.02 05.2017 5Operation Sirona Dental A/S 5.2Using the menu and display information Operating Instructions DAC Universal 5.2.3 Default settings The default settings are predetermined by the manufacturer: Rest…

[Page 63] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 63 Sirona Dental A/S 5Operation Operating Instructions DAC Universal 5.3Reprocessing with the DAC Universal båÖäáëÜ 5.2.5 Drain the chamber If there is too much water left in the chamber, for in…

[Page 64] Sirona DAC UNIVERSAL

66 15 764 D3472 64 D3472.201.21.01.02 05.2017 5Operation Sirona Dental A/S 5.3Reprocessing with the DAC Universal Operating Instructions DAC Universal 5.3.2 Reprocessing with the Standard lid The Standard lid is used for reprocessing unwrappe…

[Page 65] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 65 Sirona Dental A/S 5Operation Operating Instructions DAC Universal 5.3Reprocessing with the DAC Universal båÖäáëÜ 5. Insert the Standard lid in the holder of the DAC Universal in such a way t…

[Page 66] Sirona DAC UNIVERSAL

66 15 764 D3472 66 D3472.201.21.01.02 05.2017 5Operation Sirona Dental A/S 5.3Reprocessing with the DAC Universal Operating Instructions DAC Universal  The following message appears on the display when the selected cycle is finished: &qu…

[Page 67] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 67 Sirona Dental A/S 5Operation Operating Instructions DAC Universal 5.3Reprocessing with the DAC Universal båÖäáëÜ 16. Check whether the dark status bar of the chemical indicator is in the &qu…

[Page 68] Sirona DAC UNIVERSAL

66 15 764 D3472 68 D3472.201.21.01.02 05.2017 5Operation Sirona Dental A/S 5.3Reprocessing with the DAC Universal Operating Instructions DAC Universal 5.3.3 Reprocessing with the Basket lid The Basket lid is used for reprocessing unwrapped so…

[Page 69] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 69 Sirona Dental A/S 5Operation Operating Instructions DAC Universal 5.3Reprocessing with the DAC Universal båÖäáëÜ 3. Open the snap lock on the basket and take the small lid off the basket. 4. …

[Page 70] Sirona DAC UNIVERSAL

66 15 764 D3472 70 D3472.201.21.01.02 05.2017 5Operation Sirona Dental A/S 5.3Reprocessing with the DAC Universal Operating Instructions DAC Universal 6. Reattach the wire lid on the basket. 7. Close the basket. 8. Insert the Basket lid in th…

[Page 71] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 71 Sirona Dental A/S 5Operation Operating Instructions DAC Universal 5.3Reprocessing with the DAC Universal båÖäáëÜ 9. Press the «Start»-Key.  A prompt to select the lid appears on …

[Page 72] Sirona DAC UNIVERSAL

66 15 764 D3472 72 D3472.201.21.01.02 05.2017 5Operation Sirona Dental A/S 5.3Reprocessing with the DAC Universal Operating Instructions DAC Universal 15. Remove the Basket lid from the holder on the DAC Universal at the end of the cycle to …

[Page 73] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 73 Sirona Dental A/S 5Operation Operating Instructions DAC Universal 5.3Reprocessing with the DAC Universal båÖäáëÜ 5.3.4 Reprocessing with the Flex lid The Flex lid is used for cleaning and hig…

[Page 74] Sirona DAC UNIVERSAL

66 15 764 D3472 74 D3472.201.21.01.02 05.2017 5Operation Sirona Dental A/S 5.3Reprocessing with the DAC Universal Operating Instructions DAC Universal  There are instruments on the Flex lid; the DAC Universal is still open and ready for a…

[Page 75] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 75 Sirona Dental A/S 5Operation Operating Instructions DAC Universal 5.3Reprocessing with the DAC Universal båÖäáëÜ 11. Press the «Clear»-Key.  The Flex lid will open completely. 12…

[Page 76] Sirona DAC UNIVERSAL

66 15 764 D3472 76 D3472.201.21.01.02 05.2017 5Operation Sirona Dental A/S 5.3Reprocessing with the DAC Universal Operating Instructions DAC Universal  After the end of the cycle, the DAC Universal automatically reactivates lubrication fo…

[Page 77] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 77 Sirona Dental A/S 6Cleaning and maintenance Operating Instructions DAC Universal båÖäáëÜ 6 Cleaning and maintenance After every 50 cycles, the DAC Universal will give a 2 second acoustic alarm a…

[Page 78] Sirona DAC UNIVERSAL

66 15 764 D3472 78 D3472.201.21.01.02 05.2017 6Cleaning and maintenance Sirona Dental A/S 6.1Checking oil supply to the chamber Operating Instructions DAC Universal 6.1 Checking oil supply to the chamber Frequency: weekly or every 50 cycles (w…

[Page 79] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 79 Sirona Dental A/S 6Cleaning and maintenance Operating Instructions DAC Universal 6.3Checking water supply to the lids båÖäáëÜ 6. Interrupt this process by pressing the «Clear»-Key as…

[Page 80] Sirona DAC UNIVERSAL

66 15 764 D3472 80 D3472.201.21.01.02 05.2017 6Cleaning and maintenance Sirona Dental A/S 6.4Checking oil supply to the Standard lid Operating Instructions DAC Universal 6.4 Checking oil supply to the Standard lid Frequency: weekly or every 50…

[Page 81] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 81 Sirona Dental A/S 6Cleaning and maintenance Operating Instructions DAC Universal 6.5Drain the water tank båÖäáëÜ 6.5 Drain the water tank Frequency: weekly or every 50 cycles (whichever comes …

[Page 82] Sirona DAC UNIVERSAL

66 15 764 D3472 82 D3472.201.21.01.02 05.2017 6Cleaning and maintenance Sirona Dental A/S 6.6Cleaning the water tank Operating Instructions DAC Universal  Water is pumped from the water tank through the chamber and into the drain tank / dr…

[Page 83] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 83 Sirona Dental A/S 6Cleaning and maintenance Operating Instructions DAC Universal 6.6Cleaning the water tank båÖäáëÜ 4. Lift the water tank straight out of the DAC Universal. 5. Clean the water…

[Page 84] Sirona DAC UNIVERSAL

66 15 764 D3472 84 D3472.201.21.01.02 05.2017 6Cleaning and maintenance Sirona Dental A/S 6.6Cleaning the water tank Operating Instructions DAC Universal 8. Wipe the tank dry with a clean cloth. 9. Wipe the sieve dry with a clean cloth. 10. Gr…

[Page 85] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 85 Sirona Dental A/S 6Cleaning and maintenance Operating Instructions DAC Universal 6.7Cleaning the drain filter båÖäáëÜ 7. Press the «Menu/Enter»-Key to confirm the selection. 8. Press…

[Page 86] Sirona DAC UNIVERSAL

66 15 764 D3472 86 D3472.201.21.01.02 05.2017 6Cleaning and maintenance Sirona Dental A/S 6.8Removing and cleaning the drain tank Operating Instructions DAC Universal 3. Clean the drain filter thoroughly under running water. 4. Screw the drain…

[Page 87] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 87 Sirona Dental A/S 6Cleaning and maintenance Operating Instructions DAC Universal 6.8Removing and cleaning the drain tank båÖäáëÜ 1. Unscrew the lid of the drain tank. 2. Remove the condensatio…

[Page 88] Sirona DAC UNIVERSAL

66 15 764 D3472 88 D3472.201.21.01.02 05.2017 6Cleaning and maintenance Sirona Dental A/S 6.9Checking the adapters and the respective O-rings Operating Instructions DAC Universal 6. Fill the clean drain tank with water up to the lowest mark on…

[Page 89] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 89 Sirona Dental A/S 6Cleaning and maintenance Operating Instructions DAC Universal 6.10Cleaning the empty chamber båÖäáëÜ Process ✔ The lid is at hand and cooled down. 1. Place the lid on the…

[Page 90] Sirona DAC UNIVERSAL

66 15 764 D3472 90 D3472.201.21.01.02 05.2017 6Cleaning and maintenance Sirona Dental A/S 6.10Cleaning the empty chamber Operating Instructions DAC Universal 3. Place a NitraClean tablet in the empty chamber. See «Functions with the Bask…

[Page 91] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 91 Sirona Dental A/S 6Cleaning and maintenance Operating Instructions DAC Universal 6.11Cleaning the housing of the unit båÖäáëÜ 10. Remove the Standard lid from the holder on the DAC Universal a…

[Page 92] Sirona DAC UNIVERSAL

66 15 764 D3472 92 D3472.201.21.01.02 05.2017 6Cleaning and maintenance Sirona Dental A/S 6.13Checking the safety ring Operating Instructions DAC Universal 1. Clean all available lids with warm water 2. Clean the lids with a solution of 70% is…

[Page 93] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 93 Sirona Dental A/S 6Cleaning and maintenance Operating Instructions DAC Universal 6.14Changing the drain filter båÖäáëÜ 6.14 Changing the drain filter Frequency: every 4 weeks or every 200 cy…

[Page 94] Sirona DAC UNIVERSAL

66 15 764 D3472 94 D3472.201.21.01.02 05.2017 6Cleaning and maintenance Sirona Dental A/S 6.15Changing the O-rings on the adapters Operating Instructions DAC Universal 6.15 Changing the O-rings on the adapters Frequency: every 3 months or ever…

[Page 95] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 95 Sirona Dental A/S 7Errors / Display messages Operating Instructions DAC Universal båÖäáëÜ 7 Errors / Display messages Any errors that cannot be immediately recognized by the user but must be rem…

[Page 96] Sirona DAC UNIVERSAL

66 15 764 D3472 96 D3472.201.21.01.02 05.2017 7Errors / Display messages Sirona Dental A/S 7.1Display messages and what they mean Operating Instructions DAC Universal 7.1 Display messages and what they mean The display messages have the follow…

[Page 97] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 97 Sirona Dental A/S 7Errors / Display messages Operating Instructions DAC Universal 7.1Display messages and what they mean båÖäáëÜ 48 The mainboard is defective. Please contact your authorized d…

[Page 98] Sirona DAC UNIVERSAL

66 15 764 D3472 98 D3472.201.21.01.02 05.2017 7Errors / Display messages Sirona Dental A/S 7.1Display messages and what they mean Operating Instructions DAC Universal 70 The pressure in the chamber does not drop quickly enough after the pres…

[Page 99] Sirona DAC UNIVERSAL

66 15 764 D3472 D3472.201.21.01.02 05.2017 99 Sirona Dental A/S 7Errors / Display messages Operating Instructions DAC Universal 7.1Display messages and what they mean båÖäáëÜ 83 The temperature sensor in the chamber is defective. Please …

[Page 100] Sirona DAC UNIVERSAL

66 15 764 D3472 100 D3472.201.21.01.02 05.2017 7Errors / Display messages Sirona Dental A/S 7.1Display messages and what they mean Operating Instructions DAC Universal If the same error is reported every time it is reset, the safety system has…